Troubleshooting
6-67
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0408016
Warning
Can’t detect
UPS
System can’t detect UPS.
Please keep AC plugged in to
avoid data loss. If this issue
persists. Please call for service.
Communication between
UPS and WKS is abnormal.
1. PC USB for UPS may be
abnormal
2.connection between PC
and UPS may be abnormal
3. UPS may be abnormal.
4. Reinstall WKS software
0408018
UPS needs to
be replaced for
aging of battery
UPS battery aged. Please call
service to replace the battery.
0408019
Warning
UPS battery
overloads
UPS battery overloaded. Please
restart the system to avoid
data loss. If this issue persists,
please call for service.
1, PC power supply may be
abnormal
0409501
Error
OSGI boot up
failed
Application modules boot up
failure. Please restart the
system. If this issue persists,
please call for service.
0409502
Error
C module boot
up failed
System modules boot up
failure. Please restart the
system. If this issue persists,
please call for service.
0409503
Error
CORBA
communication
failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0404501
Error
TRU module
resolve object
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0404502
Error
TRU module
Invoke corba
service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.