Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
 
7-76 
1.  Remove the image intensifier refer to the bundle cable replacement instruction. 
2.  Loosen four screws and remove the I.I plate, and then pull out from the C-arm. 
 
   
3.  Loosen the brake by enclasping the C-arm with at least 2 people’s effort and pull out the truck along 
the orbital direction, then place the C-arm with generator on the soft underlay. 
 
4.  Replace the Rolling Wheels with Retainer ring clamp. 
 
5.  Reassemble and recover the system per the reverse order.