Guidance and manufacturer’s declaration - Electromagnetic Immunity
The Brivo OEC 715/785/865 Mobile C-arm X-ray Product is intended to use in the electromagnetic environment
specified below. The purchaser or operator of the Brivo OEC 715/785/865 Mobile C-arm X-ray Product should
assure that it is used in an electromagnetic environment as described below:
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power
(W)
Distance
(m)
I
MMUNITY
TEST LEVEL
(V/m)
385
380 –390
TETRA 400
Pulse
modulation
b)
18 Hz
1,8
0,3
27
450
430 – 470
GMRS 460,
FRS 460
FM
c)
5 kHz deviation
1 kHz sine
2
0,3
28
710
704 – 787
LTE Band 13,
17
Pulse
modulation
b)
217 Hz
0,2
0,3
9
745
780
810
800 – 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2
0,3
28
870
930
1 720
1 700 –
1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation
b)
217 Hz
2
0,3
28
1 845
1 970
2 450
2 400 –
2 570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
b)
217 Hz
2
0,3
28
5 240
5 100 –
5 800
WLAN 802.11
a/n
Pulse
modulation
b)
217 Hz
0,2
0,3
9
5 500
5 785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the
ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a)
For some services, only the uplink frequencies are included.
b)
The carrier shall be modulated using a 50 % duty cycle square wave signal.
c)
As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.