Technical specifications and default settings: Specifications
2048723-002C CARESCAPE V100 Vital Signs Monitor A-5
Ohmeda SpO
2
Ohmeda SpO2 specifications
Measurement range
SpO
2
1 to 100%
Pulse rate 30 to 250 bpm
Perfusion Index Value (Plr) 0.00 to 9.99
Accuracy*
Saturation
Adult* 70 to 100% ±2 digits, whichever is greater,
(without motion)
Neonate* 70 to 100% ±3 digits (without motion)
Adult/Neonate** 70 to 100% ±3 digits (during clinical motion)
Low perfusion 70 to 100% ±2 digits (during clinical low
perfusion)
Pulse rate
Adult /Neonate 30 to 250 bpm: ± 2 digits or ± 2%, whichever is
greater, (without motion)
30 to 250 bpm: ± 5 digits (during motion)
Low perfusion 30 to 250 bpm: ± 3 digits
*SpO
2
measurement accuracy is based on deep hypoxia studies using OxyTip++ sensors on healthy adult volunteer subjects. Arterial blood
samples were analyzed simultaneously on multiple CO-oximeters. This variation equals plus or minus one standard deviation. Plus or minus one
standard deviation encompasses 68% of the population.
**Applicability: TS-AF sensors.
NOTE
Accuracy may vary for some sensors; always check the instructions for the sensor.