4-10 CARESCAPE V100 Vital Signs Monitor 2048723-002C
Maintenance: Long-term storage
2. Loosen the single screw at the bottom, on the back of the scanner, and
remove the battery cover.
3. Disconnect the old battery and replace with a new, high quality 9-volt
alkaline battery in the same location.
4. Replace the battery cover, and tighten the screw.
5. Reconnect the scanner cable to the host communication port, and tighten
the two thumbscrews.
Fuses
The monitor contains three fuses. The fuses are mounted within the monitor. The
fuses protect the low voltage DC input, the main battery, and the remote alarm
output. The +5 V output on the host port connector is regulated by internal
supply. Fuses are not replaceable.
Cover
9-volt battery
Single screw