A-8 CARESCAPE V100 Vital Signs Monitor 2048723-002C
Technical specifications and default settings: Specifications
Specifications: Nellcor sensor accuracy
NOTE
All Nellcor O
XIMAX
sensors must be used with the NELL cable; the SCP-10 cable. RS-10 and Oxisensor II sensors are
not compatible with the CARESCAPE V100 vital signs monitor.
Sensor model
SpO
2
range
70% to 100%
O
XIMAX
MAX-A, MAX-AL ± 2 digits
MAX-N
(adult)
± 2 digits
MAX-N
*
(neonate)
± 3 digits
MAX-P ± 2 digits
MAX-I ± 2 digits
MAX-FAST ± 2 digits
SC-A (adult) ± 2 digits
SC-PR (neonate) ± 3 digits
SC-NEO ± 3 digits
MAX-R
**
± 3.5 digits
OxiCliq
OxiCliq A ± 2.5 digits
OxiCliq P ± 2.5 digits
OxiCliq N
(adult)
± 2.5 digits
OxiCliq N
*
(neonate)
± 3.5 digits
OxiCliq I ± 2.5 digits
Reusable sensor models
D-YS (infant to adult) ± 3 digits
D-YS (neonate) ± 4 digits