3-8 CARESCAPE V100 Vital Signs Monitor 2048723-002C
Installation: Configuring the monitor
and select the parameter function. To change alarm settings, press the 
Alarms button.
Inflation pressure default setting
Procedures 
1. Enter the configuration mode: with the monitor off, press and hold the Menu 
button at the same time as pressing and holding the On/Off button for 3 
seconds.
2. Use the +/- buttons to increment or decrement the inflate pressure default 
setting.
3. To exit the configuration mode, turn the unit off. To continue with additional 
configuration settings, press Menu.
Vital sign alarm default settings
Procedures
1. Enter the configuration mode: with the monitor off, press and hold the Menu 
button at the same time as pressing and holding the On/Off button for 3 
seconds. After the unit enters the configuration mode, press Alarms. At any 
point in the configuration mode menu, Alarms default can be selected. 
2. To set or change the default setting, press the Alarms button to select alarm 
setting. Use the +/- buttons to increment or decrement the individual 
settings. 
NOTE
For the Alarms default setting to be saved, you must advance the menu 
through the SpO
2
 settings. 
3. To exit the configuration mode, turn the unit off. To continue with additional 
configuration settings, press Menu.
Reverting to the factory default vital sign alarm limits
WARNING
The Line Frequency mode (for Datex-Ohmeda oximetry) must 
be set according to each country’s electrical power utilities 
implementation; and it must be checked and reset any time the 
monitor is set to or reverts to factory default settings.
To revert to the factory default vital sign alarm limit settings, the monitor must 
be disconnected from the DC power supply and from the monitor battery. Refer 
to “Replacing the main battery on the monitor” on page 4-8 for DC power supply 
and battery disconnection/reconnection instructions.
When reverting to factory default settings, the user settings (including alarm 
limits and inflation pressure), date/time, and the Ohmeda TruSignal SpO
2
 Line 
Frequency mode (LF) will go back to default values. Refer to “Configuration mode 
settings” on page 3-5 to configure the factory default user settings.