Temperature
Number of channels 2
Input specifications
Probe type YSI Series 400 or 700 (determined by input cable)
Temperature range 0°C to 45°C (32°F to 113°F)
Resolution ± 0.1°C
Output specifications
Parameters displayed T1, T2
Accuracy
(independent of source) ± 0.1°C for YSI series 400; ± 0.3°C for YSI series 700 probes
Alarms User-selectable upper and lower limits for T1, T2
Cardiac output
Input specifications
Probe type In-line or bath probe
Catheter size 5F, 6F, 7F, 7.5F and 8F
Injectate volume 3, 5 or 10 cc
Output specifications
Parameters displayed Cardiac output, blood temperature, injectate temperature and trial number
Range
Cardiac output 0.2-15 (liters per minute)
Blood temperature 30-42°C
Injectate temperature 0-30°C
Accuracy
Cardiac output ± 5%
Blood temperature ± 0.2°C
Injectate temperature ± 0.3°C
Frequency response dc to 15 Hz ± 2 Hz
Pulse oximetry
Parameters monitored Arterial oxygen saturation (SpO
2
) and peripheral pulse rate
(PPR)
SpO
2
range Nellcor 1-100%; Masimo 30-100%; GE Ohmeda 30-100%
PPR range Nellcor 20-300 BPM; Masimo 25-240 BPM; GE Ohmeda 30-250
Accuracy Actual accuracy depends on probe. Please reference
manufacturer’s specifications.
Nellcor SpO
2
± 2 digits (70-100% SpO
2
)
Masimo SpO
2
± 2% Adults/Pediatric (70-100% SpO
2
)
GE Ohmeda SpO
2
± 2% (70-100% SpO
2
), SpO
2
± 3% Neonates, ≤ 69% unspecified
PPR ± 3 beats per minute
Alarms
User-selectable upper and lower limits for SpO
2
and PPR