Chapter 2. Error Codes
16 DigitalFlow™ XGM868i Service Manual
E9: Pressure In
Problem: This message indicates a pressure input error.
Cause: The pressure exceeds the specified limits for the analog inputs option card or no input device is connected.
Action: Check the pressure transmitter and the connecting cable. Refer to Chapter 1, Calibration, and recalibrate
the analog inputs option card.
E10: Special Input
Problem: This message indicates a special input error.
Cause: The special input exceeds the specified limits for the analog inputs option card.
Action: Check the special input device and the connecting cable. Refer to Chapter 1, Calibration, and recalibrate
the analog inputs option card.
E11: Not Used
This error code message is not currently in use.
E12: Low Pressure
Problem: This error code is triggered when the pressure measurement is below the low pressure switch setting.
Cause: The pressure measurement exceeds the specified limits as set in the Channelx-Input/Output submenu.
Cause: Check the limit value (0 to 5,000 psia) entered into the Channelx-Input/Output submenu, as described in
Chapter 1, Programming Site Data, of the Programming Manual.
E13: Over Range
Problem: This error code message indicates that the present measurement exceeds the range of the meter.
Cause: A internal mathematical overflow has occurred in either the volumetric or mass flow calculations.
Action: Select larger measurement units or a shorter time interval for the current measurement parameter. For
example, choose KSCF/M instead of SCF/M in the Channelx-System menu. See Chapter 2, Initial Setup, of
the Startup Guide for instructions.
E14: Totals Overflow
Problem: The totalizers are unable to keep up with the total accumulated flow signals.
Cause: The programmed units/pulse value is too small.
Action: Select a larger number of units/pulse value.