Appendix B. Calibration and Testing with PanaView
54 DigitalFlow™ XGM868i Service Manual
B.5.2 Uploading XGM868i Memory
Note: This option is for factory use only.
The Upload XGM Memory option is another diagnostic tool.
IMPORTANT: Consult GE before using this data to change any parameters.
1. In the Calibration/Test menu, double-click on the Upload XGM Memory option.
2. The first prompt shows the Hex Start Address. Click [Next Item].
3. The next prompt displays the value for the Hex # Bytes. Click [Next Item].
4. PanaView now offers two options for Uploading the XGM memory.
If you double-click on To Display, the right pane of the window appears similar to Figure 14, and the data is
available for immediate evaluation. Click [Exit Page] to close the window and return to the Calibration/Test
menu.
Figure 14: XGM868i Memory Data Displayed on PanaView
If you double-click on To File, the Save As window (similar to Figure 13 on page 53) opens. Enter a file name
and click [Save]. PanaView stores the array data in a Windows file, and closes the Calibration/Test window.