g 
Product Data Sheet – rev 4.2 
July 4th 2003  
Page 7 
 
 
 
LightSpeed
  
With Xtream 
CT Scanner System 
GE Medical Systems-America 
Milwaukee, USA  - Fax: 1 414 544 3384 
GE Medical Systems-Asia: 
Tokyo, Japan - Fax: 81 425 85 5490 
Hong Kong - Fax: 852 2559 3588 
GE Medical Systems-Europe 
•  0.584 mm limiting resolution 
•  Nominal Thickness: 0.625, 1.25 mm 
Axial Scans 
•  4.0 lp/cm @ 50% MTF 
Axial Scan Parameters 
•  6.5 lp/cm @ 10% MTF 
Multi-slice acquisitions and short interscan 
delays significantly reduce potential missed 
registration between scans by increasing the 
number of scans possible in a patient breath 
hold.  Contrast agents may be better utilized 
as well due to significantly faster scans. 
•  8.5 lp/cm @ 0% MTF 
Scan Time: 
•  0.5, 0.6, 0.7, 0.8, 0.9, 1.0, 2.0, 3.0 and 4.0 
second full scans (360° acquisition) 
Hi-Res Algorithm (Edge).
 
•  0.324 mm limiting resolution 
Scan Technique: 
Axial Multi-slice Prescription 
•  8.5 lp/cm @ 50% MTF 
•  kVp:   80, 100, 120, 140 
•  mA: 10 to 440, in 5 mA increments 
•  13.0 lp/cm @ 10% MTF 
Simplified scan prescriptions and easy-to-use 
default protocols make the Lightspeed
16
 CT 
Scanner fast and efficient in patient set-up.  
Axial protocols are nearly identical to helical 
scan protocols. 
•  Power:  0.8 to 53.2 kW  
•  15.4 lp/cm @ 0% MTF 
•  Focal Spot Selection: 
Line pair values decrease with larger focal 
spot (by 5% with Standard and by 7% with 
Hi-Res); limiting resolution is unaffected. 
•  Small spot for up to 24 kW 
•  Large spot for greater than 24 kW 
Axial Multi-slice Modes 
Scan Plane Geometry: 
Low-Contrast Detectability 
 
•  +/- 30° gantry tilt, in  0.5° increments 
The  LightSpeed
16 
 acquires axial scans in 
sets of 8 or 16 contiguous images in one 360° 
rotation.  
 
•  Longitudinal positioning in 0.01 mm per slice 
increment. Gantry display in 0.5 mm 
increments. 
On 8 inch (20 cm) CATPHAN phantom:  
5 mm @ .3% at 13.3 mGy  
3 mm @ .3% at 37.2 mGy  
For each rotation of the gantry, the  
LightSpeed
16 
 collects 16 rows of scan 
data. There are five reconstruction modes 
available for creating images from the 
multi-slice scan data (1i, 2i, 4i, 8i, and 16i).  
By using 1i, 2i, 4i, and 8i reconstruction 
modes, scan data can be combined prior 
to image reconstruction to create slices 
with reduced partial-volume artifacts.  This 
is particularly useful for posterior-fossa 
imaging. 
Interscan Delay (ISD):  
Noise:
 
•  Minimum ISD with table movements of  
0 - 10 mm: 1.0 sec. 
On either an AAPM water phantom or GE 
Quality Assurance phantom: 
•  Minimum ISD with table movements of more 
than 10 mm and up to 20 mm: 1.3 sec 
0.32%  +/- 0.03% at 28.5 mGy (2.85 Rad) 
•  User-selectable. 
 
Inter Group Delay (IGD): 
CTDI:
 
•  Minimum IGD is the same as minimum ISD; 
also user-selectable. 
 
On CTDI Head and Body Dose Reference 
Phantoms: 
1i Mode: 
Scan-to-Scan Cycle: 
CTDI
vol
 expressed in mGy/100 mAs for 
IEC pitch 1: (normalized to a pitch of 1). 
•  Produces 1 image per rotation 
•  Minimum scan-to-scan cycle of 1.5 seconds 
possible for 0.5 sec. scan speed with 
minimum ISD's. 
•  Nominal Thickness: 1.25, 5, 10 mm 
Head  20.9 mGy/100 mAs 
Body  10.6 mGy/100 mAs 
2i Mode: 
Scan Fields of View: 
•  Produces 2 images per rotation 
•  25 cm for adult head 
•  Nominal Thickness: 0.625, 2.5, 5, 7.5, 10 
mm 
•  25, 50 cm for body 
•  25 cm for pediatric head 
4i Mode: 
Scan with no table incrementation, 
contiguous image location, or skipped 
image location are possible. Overlapped 
axial scans are not possible. 
•  Produces 4 images per rotation 
•  Nominal Thickness: 2.5, 3.75, 5 mm  
8i Mode: 
•  Produces 8 images per rotation 
Axial Image Reconstruction 
•  Nominal Thickness: 1.25, 2.5 mm 
Reconstruction Algorithms: Soft Tissue, 
Standard, Detail, Bone, Bone Plus, Lung and 
Edge. 
16i Mode: 
•  Produces 16 images per rotation 
* Option