LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 43
Declaration of Immunity
This system is suitable for use in the following environment. The user must assure that the system is used according to the specified guidance and only in the
electromagnetic environment listed.
Table 11: Declaration of Immunity
Immunity Type Test Level Compliance EMC Environment and Guidance
IEC 61000-4-2
Static discharge
(ESD)
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.
Mains power quality should be that of a typical commercial
and/or hospital environment. If the user requires continued
operation during power mains interruptions, it is
recommended that the system be powered from a UPS or a
battery.
NOTE: UT is the a.c. mains voltage prior to application of the
test level.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial and/
or hospital environment.
Separation distance to radio communication equipment must
be maintained according to the method below. Interference
may occur in the vicinity of equipment marked with the
symbol:
IEC 61000-4-4
Electrical fast
transient/burst
± 2 kV for mains
± 1 kV for SIP/SOP
± 2 kV for mains
± 1 kV for SIP/SOP
IEC 61000-4-5 Surge Immunity ± 1 kV differential
± 2 kV common
± 1 kV differential
± 2 kV common
IEC 61000-4-11
Voltage dips, short interruptions and
voltage variations on mains supply
< 5% U
T
(> 95% dip) for 0.5 cycle;
40% U
T
(60% dip) for 5 cycles;
70% U
T
(30% dip) for 25 cycles;
< 5% U
T
(>95% dip) for 5 sec
< 5% U
T
(> 95% dip) for 0.5 cycle;
40% U
T
(60% dip) for 5 cycles;
70% U
T
(30% dip) for 25 cycles;
< 5% U
T
(>95% dip) for 5 sec
IEC 61000-4-8
Power frequency (50/60 Hz) magnetic
field
3 A/m 3 A/m
IEC 61000-4-6
Conducted RF
3 V
RMS
150 kHz - 80 MHz
3 V
RMS
150 kHz - 80 MHz
IEC 61000-4-3
Radiated RF
3 V/m
80 MHz - 2.5 GHz
3 V/m
80 MHz - 2.5 GHz
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.