LOGIQ
LOGIQ E10 Series: Power Assistant Battery Option
Battery Icon on the back of
the monitor depicts Power
Assistant is available on
the system.
DO NOT perform the following while in PA
Mode:
ü Connect and disconnect probes
ü Disconnect peripheral devices, such as the
Volume Navigation unit
ü Eject or insert CD/DVD media or external
USB storage devices
To Enter Power Assistant (PA)- Remove the
power cord from AC power while the system is on.
Before unplugging the power cord and starting
PA, if you are performing any of the following
actions, ensure that they are complete:
Any export, input, or “Save As” to external
media, such as CD, DVD, USB
DICOM
Retrieve from Query/Retrieve
DVR recording should be stopped
Exit Volume Navigation by Exit and Clear if
you are using Volume Navigation
PA
Flashing lights indicate the battery level is LOW
Flashing operator panel lights are meant to notify
the user that the battery charge level is low. They
will begin to flash when the charge level is at 40%
and lower. The user should consider this a visual
clue to plug into a power source as soon as
possible.
If the charge level of the battery falls below 20%,
the operator panel lights will increase the
frequency of flashing and the system will prepare
for shutdown.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E10 Series and is the answer not in the manual?

GE LOGIQ E10 Series Specifications

General IconGeneral
ModelLOGIQ E10 Series
XDclear TechnologyYes
Electronic 4DYes
Power Requirements100-240 V AC, 50/60 Hz
CategoryUltrasound Imaging System
Touch Panel12.1-inch touch panel
StorageIntegrated hard drive
ConnectivityDICOM, USB, Ethernet
ApplicationAbdominal, Vascular, Cardiac, Musculoskeletal, Small Parts, OB/GYN
eM6C probeAvailable
Transducer Connectors4 active transducer ports
Probe Frequency Range1-20 MHz
ArchitecturecSound™ Architecture