LOGIQ
LOGIQ E10 Series: Power Assistant Battery Option
Battery Icon on the back of
the monitor depicts Power
Assistant is available on
the system.
DO NOT perform the following while in PA
Mode:
ü Connect and disconnect probes
ü Disconnect peripheral devices, such as the
Volume Navigation unit
ü Eject or insert CD/DVD media or external
USB storage devices
To Enter Power Assistant (PA)- Remove the
power cord from AC power while the system is on.
Before unplugging the power cord and starting
PA, if you are performing any of the following
actions, ensure that they are complete:
• Any export, input, or “Save As” to external
media, such as CD, DVD, USB
• DICOM
Ⓡ
Retrieve from Query/Retrieve
• DVR recording should be stopped
• Exit Volume Navigation by Exit and Clear if
you are using Volume Navigation
PA
Flashing lights indicate the battery level is LOW
• Flashing operator panel lights are meant to notify
the user that the battery charge level is low. They
will begin to flash when the charge level is at 40%
and lower. The user should consider this a visual
clue to plug into a power source as soon as
possible.
• If the charge level of the battery falls below 20%,
the operator panel lights will increase the
frequency of flashing and the system will prepare
for shutdown.