Site Preparations
3-6 Ultrasound SystemCommon Service Information
Direction 5444964-100 English
Rev. 5
Electrical requirements for Console Ultrasound Systems
Contents:
• ‘Electrical Requirements - Invenia ABUS’ on page 3-6
• ‘Electrical requirements - LOGIQ Consoles’ on page 3-7
• ‘Electrical Specifications for LOGIQ C Series and LOGIQ
C3/C5 Premium’ on page 3-8
• ‘Electrical Requirements - Vivid Consoles’ on page 3-9
• ‘Electrical Requirements - Voluson Consoles’ on page 3-10
Table 3-2: Electrical Requirements - Invenia ABUS
Product Voltage Tolerance
Current or Power
Consumption
Frequency
Invenia ABUS
Scan Station
100 - 240 VAC ±10% 10 A 50/60 Hz