EasyManuals Logo

GE LOGIQ P5 Technical Publication

GE LOGIQ P5
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #63 background imageLoading...
Page #63 background image
General Ultrasound system requirements
Ultrasound SystemCommon Service Information 3-11
Direction 5444964-100 English Rev. 5
Electrical requirements for Laptop Ultrasound Systems
Contents:
• ‘Electrical Requirements - LOGIQ Laptops’ on page 3-11
‘Electrical Requirements - Vivid Laptops’ on page 3-12
‘Electrical Requirements - Voluson Laptops’ on page 3-12
‘Docking Cart for LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid| e/
Vivid iq Power Requirements’ on page 3-13
• ‘Isolation Cart for LOGIQBook XP Series/LOGIQ e/
LOGIQ e Vet/LOGIQ i/Vivid e’ on page 3-14
Table 3-7: Electrical Requirements - LOGIQ Laptops
Product Voltage Tolerance
Current or
Power
Consumption
Frequency
LBAC-66 adapter for
LOGIQ Book
100-240 VAC ±10%
120 VA 1.2-0.5 A
50-60 Hz
GE-90W adapter for
LOGIQ Book
108 VA 1.08-0.45
A
ADM-9020M-GE for
LOGIQ Book XP Series
Adapter for LOGIQ e and
LOGIQ e Vet R7.x.x and
below.
130 VA max. 1.3 A
(max.) 50/60 Hz
Adapter for LOGIQ i
Adapter for LOGIQ e
R8.x.x, R9.x.x
160 VA max. 1.6 A
(max.)
AC Adapter for LOGIQ
V2/V1
2 A 50-60 Hz

Table of Contents

Other manuals for GE LOGIQ P5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 and is the answer not in the manual?

GE LOGIQ P5 Specifications

General IconGeneral
Power Supply100-240VAC, 50/60Hz
Display15" LCD
Doppler ModesColor Doppler, Power Doppler
Imaging ModesB-Mode, M-Mode
ProbesConvex, Linear, Phased Array
ApplicationsAbdominal, Vascular, Cardiac, OB/GYN, Small Parts, Musculoskeletal
ConnectivityDICOM, USB, Ethernet
Probe Connectors3
PortabilityMobile
Operating Temperature10°C to 40°C
Relative Humidity30%

Related product manuals