IRECTION 5394227, 12   LOGIQ S8/LOGIQ E8 SERVICE MANUAL 
3 - 72 Section 3-12 - Configuring Connectivity
Section 3-12
Configuring Connectivity
3-12-1 Overview
You use Connectivity functionality to set up the connection and communication protocols for the 
ultrasound system. This page gives an overview of each of the Connectivity functions. Each function is 
described in detail in the following pages. 
3-12-2 Contents in this section
3-12-1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-72
3-12-2 Contents in this section. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-72
3-12-3 Structured Reporting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-72
3-12-4 Connectivity Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-73
3-12-5 TCPIP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-74
3-12-6 Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-76
3-12-7 Service  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-77
3-12-3 Structured Reporting
DICOM Structured Reporting provides the results of a procedure as structured data elements (well-
defined fields) as opposed to unstructured data (large amounts of text undifferentiated by individual 
fields). This greatly improves query capability. DICOM Structured Reporting creates coded clinical data 
that can be used for clinical research, outcomes analysis, and disease management.
DICOM Structured Reporting is a standardized format for medical results. LOGIQ S8 supports the 
following templates:
• OB-GYN REPORT TEMPLATES
• VASCULAR ULTRASOUND REPORT TEMPLATES
• The system supports selectable vessels with their locations for B-Mode measurements.
• For Vascular studies (Bypass Graft, UEV Map, LEV Map, LEA/UEA, and Carotid), additional 
measurement locations were added. For instance, for a Bypass Graft, there are now measurements 
at Inflow, Anast, Thigh, Knee, Calf, Ankle, Graft, RunOff, Pre-Sent, Stent, Post-Stent, and Outflow 
locations.
• ECHOCARDIOGRAPHY PROCEDURE REPORT TEMPLATES
These templates do not support all M&A results for the LOGIQ S8.
3-12-3-1 Supported parameters
The DICOM supported parameters are listed in the DICOM Conformance Statement at the following 
web site under DICOM - Ultrasound:
http://www.gehealthcare.com/usen/interoperability/dicom/