9 - 10 Section 9-6 - System maintenance
9-6-3 Cleaning
9-6-3-1 General Cleaning
Frequent and diligent cleaning of the LOGIQ™ S8 reduces the risk of spreading infection from person 
to person, and also helps to maintain a clean work environment. 
9-6-3-2 Appropriate Cleaning/Disinfectant Agents
OK = Available agent
N/A = Not Available
Table 9-7    General Cleaning
Step Item Description
1
Console
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system. 
Be careful not to get the cloth too wet so that moisture does not enter the console.
Caution: DO NOT allow any liquid to drip or seep into the system.
2
Monitor
Clean the monitor surface with a fluid detergent in warm water on a soft, damp cloth. 
Caution: DO NOT spray any liquid directly onto the LOGIQ™ S8 covers, monitor, keyboard, etc.
Table 9-8:  Appropriate Cleaning/Disinfectant Agents
Cleaning/
Disinfectant 
Agent
Monitor (Glass)
Monitor Frame 
(Panel)
System 
Cabinet
Measurement 
Selection Menu Operator Controls
Mild, Non-
Abrasive Soap 
and Water
OK OK OK OK OK
Ammonia N/A OK N/A N/A N/A
<Disinfectant>
Bleach (10 to 1 
Ratio of 5% 
Home Bleach)
N/A OK N/A N/A OK
<Cleaner/
disinfectant>
Hydrogen 
Peroxide / 
Hydrogen 
Peroxide Wipes
N/A OK N/A N/A N/A
Other 
recommended 
Cleaner/
Disinfectants
N/A N/A N/A N/A PDI - Super Sani Cloth
Clorox - Multi Surface 
Wipes
Hartmann - Kohrsolin Extra
Hartmann - Kohrsolin FF
Notes Never use thinner, benzene, 
alcohol (ethanol, methanol, 
or isopropyl alcohol), 
abrasive cleaners, or other 
string solvents, as these may 
cause damage to the monitor. 
Cloth should 
be damp, not 
dripping wet. 
Cloth should be 
damp, not dripping 
wet. 
DO NOT USE: 
Any cleaning/ disinfecting 
solution BESIDES 
recommended cleaner.