GE HEALTHCARE OPTIMA XR200AMX/XR220AMX
D
IRECTION 5336112-1EN, REVISION 3PRE-INSTALLATION MANUAL
Page 30 Section 4.1 Power requirements
4.1.2 Power consumption
Continuous and peak power consumption during use are not applicable for this system, because
power for use is drawn from the batteries (even if the system is connected to wall power).
Power consumption during battery charging (assuming the system is not being used during
charging) is as follows:
Draw is less than 1200 W from 100 VAC (50/60 Hz) power outlet.
Draw is less than 1300 W from 120 or 220/240 VAC (50/60 Hz) power outlet.
Typical charge time from empty to full is four hours. Charge time varies with the age of the batteries,
and whether the system is in use during charging.