0)
345- 85
/ ! ; ( . ( ( , ;,
5 HH ! " (
Note: In AOP mode and following preparation (anode rotation), exposure is made in two
stages:
D A test exposure of 15 ms allows selection of the optimum kV, focal track and filter
combination in the selected priority, and prediction of the exposure duration for
optimization of anti–scatter grid movement.
A pause during which the kV, focal track and filter are put in place.
D An exposure during which the exposure is made.
Note: In AEC mode and following preparation (anode rotation), exposure is made in two
stages:
D A test exposure of 15 ms allows prediction of the exposure duration for
optimization of anti–scatter grid movement.
D An exposure during which the exposure is made.
.
The operator must maintain pressure on the button until the buzzer
stops to get proper exposure.
.4- - 85
/ ( .! ( ( ,
( ! ((# 8( !#!
Note: If autodecompression is selected in the set–up menu, compression releases
automatically at the end of the exposure.