Revision H TONOPORT V 3-1
2001589-023
3 Functional Description
Measurement limits
- Max. cuff inflation time = 100 seconds
- Max. measurement time after start of measurement = 180 seconds
Button / Switch Display Button Display Explanation
ON all segments selftest
A100 A=accu
(rechargeable
battery)
100=100%
clock
Info H 1 Start/Start LLLL delete record
Info H 2 Start YYYY
MM
DD
hour
min
date
time
Info H 3 Start/Start
Info
LLLL P1
P2
P3
internal programs
Info H 4 Start 50 pressure offset
Start 0 calibration mode
(manometer)
Info H 5 Start 20:01 20:= FW2.0
01=German
02=English
03=French
04=Italian
05=Spanish
Info H 6 Start AAAA accu
Info bbbb battery
Info H 7 Start 1111 signal tone ON
(TONOPORT V,
V2.0 only)
Info 0000 signal tone OFF
(TONOPORT V,
V2.0 only)