TuffSat User’s Guide and Service Manual
3-2
Cleaning the oximeter
CAUTION: Follow these guidelines when cleaning the oximeter:
• Do not autoclave, pressure sterilize, or gas sterilize the oximeter.
• Use cleaning solution sparingly. Do not immerse the oximeter in liquid.
Excessive solution may flow into the oximeter and damage internal
components.
• When cleaning the display lens, do not use abrasive cleaning compounds or
any material that could damage the lens.
• Do not use petroleum-based solutions or solutions containing acetone, freon, or
harsh solvents to clean the oximeter. These substances may damage the
oximeter and cause a malfunction.
Cleaning agents
Mild detergent solution 3.4% glutaraldehyde (Cidex® Plus)
Quaternary germicides (Virex®) 70% isopropyl alcohol
1.6% phenol (Sporicidin®) 0.5% sodium hypochlorite (bleach)
1. Power off the oximeter.
2. Wipe the display lens with a cotton swab moistened with a cleaning agent.
3. Wipe the oximeter case with a soft cloth dampened with a cleaning agent. Do
not allow excess liquid to enter the sensor connector.
Sensors
To clean a durable sensor, refer to the instructions for the sensor.
CAUTION: Disposable sensors are intended for single-patient-use only.