Spectrophotometer Health &
Safety Document including
General Operating Instructions
Original Instructions

Questions and Answers

  • C
    Crystal ReedAug 18, 2025
    What to do if GE Ultrospec 2100 shows a flashing Absorbance reading of 2.00 Abs?
    • Z
      zweeksAug 19, 2025
      If you observe a flashing Absorbance reading of 2.00 Abs, it indicates that the sample needs to be diluted.
  • J
    Jason TylerAug 21, 2025
    What to do if GE Ultrospec 2100 shows Error code 001 symbol?
    • M
      Mathew StephensonAug 21, 2025
      If Error code 001 appears, it means the lamp needs changing.
  • A
    andersonaustinAug 23, 2025
    Why does GE Ultrospec 2100 switch off after calibration?
    • K
      Kathryn BerryAug 23, 2025
      The instrument may be switching off after calibration because the ON/OFF button is being held for too long. This can cause the instrument to receive both ON and OFF signals. Try adjusting the timing of your finger press at switch on.
  • J
    James MeltonAug 26, 2025
    How to fix GE Measuring Instruments failing start up calibration check?
    • N
      Nancy MoralesAug 26, 2025
      If the instrument fails the start up calibration check, ensure that the sample and reference beams are clear when the instrument is switched on.
  • J
    John CarterAug 28, 2025
    What to do if GE Ultrospec 2100 shows X against Lamp on initialisation screen?
    • M
      Michael MorganAug 28, 2025
      If an 'X' appears against the lamp on the initialization screen, it indicates that the lamp needs to be replaced.
  • J
    Jesse DouglasAug 30, 2025
    What to do if GE Ultrospec 2100 shows Absorbance readings unstable?
    • B
      brandonwilliamsAug 30, 2025
      If you are experiencing unstable absorbance readings, ensure that Reference QA is switched on, background correction is enabled, and that a 2 µl sample is being used.
  • A
    Alicia JohnsonSep 1, 2025
    Why does GE Measuring Instruments show a negative reading?
    • R
      rachael02Sep 1, 2025
      Negative readings can occur if the Reference and Test cuvettes are mixed up.
  • C
    Christopher LawsonSep 4, 2025
    What to do if GE Ultrospec 2100 instrument will not zero?
    • J
      Jesse ScottSep 4, 2025
      The instrument may not zero if the cuvette is placed in the wrong orientation or if the wrong cuvette material is used for the wavelength.
  • J
    Jonathan DavisSep 6, 2025
    What to do if GE Ultrospec 2100 Measuring Instruments shows REF is displayed when T is pressed?
    • D
      dominique91Sep 6, 2025
      If 'REF' is displayed when 'T' is pressed, it means the instrument was not zeroed before a sample measurement was attempted.
  • S
    Steven MarshallSep 8, 2025
    What to do if GE Ultrospec 2100 Measuring Instruments gives Poor reproducibility on DNA concentration and purity measurements?
    • J
      Joseph WilliamsSep 8, 2025
      Poor reproducibility in DNA concentration and purity measurements can occur if the wrong cuvette is used (not a 15 mm beam height or wrong material not letting UV light pass through), if background correction mode is not switched on, or if the measured absorbance is too low (should be >0.1 A for accurate results).

Summary

Health and Safety

1.3. Safety Notices

Explains the meaning of warning and caution symbols used in the manual for safe operation.

General Safety Information

1.4. General Safety

Lists relevant safety directives and standards the equipment complies with for safe operation.

General Safety Warnings

Provides crucial warnings regarding biohazards, chemicals, ventilation, and personal protective equipment.

Instrument Hazards and Setup

1.5. General Hazards

Details potential hazards like UV sources, high voltages, and PC connectivity safety.

Operating and Maintenance

2.6. Novaspec III;Plus Lamp Change

Step-by-step guide for replacing the lamp in Novaspec III/Plus instruments.

Troubleshooting Guide

3.1. Emergency Procedures

Outlines steps to take in case of malfunction or hazardous material spills.

GE Ultrospec 2100 Specifications