ActionStep
Insert the pipette into the probe sheath assembly through the threaded hole
in the probe sheath assembly end plug.
3
Inject the water into the probe sheath assembly.
Note:
The inserted water provides the steam inside the probe sheath assembly
during autoclaving.
4
Remove the pipette.5
Verify that the O-rings are present on the probe.6
Insert the probe into the probe sheath assembly through the threaded hole
in the probe sheath assembly end plug.
7
166 Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
6 Preparation
6.4 Prepare and install the probes
6.4.3 Autoclave probe sheath assembly