GenadyneBiotechnologies|EN(V5.0909.1) 24
THEGenadyneXLR8isintendedforuseintheelectromagneticenvironmentspecifiedbelow.Thecustomerortheuser
oftheGenadyneXLR8shouldassurethatitisusedinsuchanenvironment.
Immunitytest IEC60601
Testlevel
Compliance
level
Electromagnetic
environment–guidance
Electrostaticdischarge
(ESD)
IEC61000
‐4‐2
+/‐6kVcontact
+/‐8kVair
Passed Floorsshouldbewood,
concreteorceramictile.If
floorsarecoveredwith
syntheticmaterial,the
relativehumidityshouldbe
atleast30%.
Electricalfasttransient
/burst
IEC61000‐4‐4
+/‐2kVforpowersupply
lines
+/‐1kVforinput/output
lines
Below
Maximum
permissible
limit
Mainspowerqualityshould
bethatofatypical
commercialorhospital
environment.
Surge
IEC61000‐4‐5
+/‐1kVline(s)toline(s)
+/‐2kVline(s)toearth
Acceptable
Performance
Mainspowerqualityshould
bethat
ofatypical
commercialorhospital
environment.
Voltagedips,short
interruptionsand
voltagevariationson
powersupplyinput
lines
IEC61000‐4‐11
<5%Ut
(>95%dipinUt)for0.5
cycle
40%Ut
(60%dipinUt)for5cycles
70%Ut(30%dipinUt)for
25
cycles
<5%Ut
(>95%dipinUt)for5sec
Acceptable
Performance
Mainspowerqualityshould
bethatofatypical
commercialorhospital
environment.Iftheuserof
theGenadyneXLR8be
poweredfroman
uninterruptablepower
supplyorabattery.
Powerfrequency(50/60
Hz)magneticfield
IEC61000‐
4‐8
3A/m NonApplicable Powerfrequencymagnetic
fieldsshouldbeatlevels
characteristicsofatypical
locationinatypical
commercialorhospital
environment
NoteUtisthea.c.mainsvoltagepriortoapplicationofthetestlevel