EasyManua.ls Logo

Genadyne XLR8 - Electromagnetic Immunity Tests; Immunity Test Compliance Levels; Immunity Test Environment Guidance

Genadyne XLR8
30 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GenadyneBiotechnologies|EN(V5.0909.1) 24
THEGenadyneXLR8isintendedforuseintheelectromagneticenvironmentspecifiedbelow.Thecustomerortheuser
oftheGenadyneXLR8shouldassurethatitisusedinsuchanenvironment.
Immunitytest IEC60601
Testlevel
Compliance
level
Electromagnetic
environmentguidance
Electrostaticdischarge
(ESD)
IEC61000
42
+/‐6kVcontact
+/‐8kVair
Passed Floorsshouldbewood,
concreteorceramictile.If
floorsarecoveredwith
syntheticmaterial,the
relativehumidityshouldbe
atleast30%.
Electricalfasttransient
/burst
IEC6100044
+/‐2kVforpowersupply
lines
+/‐1kVforinput/output
lines
Below
Maximum
permissible
limit
Mainspowerqualityshould
bethatofatypical
commercialorhospital
environment.
Surge
IEC6100045
+/‐1kVline(s)toline(s)
+/‐2kVline(s)toearth
Acceptable
Performance
Mainspowerqualityshould
bethat
ofatypical
commercialorhospital
environment.
Voltagedips,short
interruptionsand
voltagevariationson
powersupplyinput
lines
IEC61000411
<5%Ut
(>95%dipinUt)for0.5
cycle
40%Ut
(60%dipinUt)for5cycles
70%Ut(30%dipinUt)for
25
cycles
<5%Ut
(>95%dipinUt)for5sec
Acceptable
Performance
Mainspowerqualityshould
bethatofatypical
commercialorhospital
environment.Iftheuserof
theGenadyneXLR8be
poweredfroman
uninterruptablepower
supplyorabattery.
Powerfrequency(50/60
Hz)magneticfield
IEC61000
48
3A/m NonApplicable Powerfrequencymagnetic
fieldsshouldbeatlevels
characteristicsofatypical
locationinatypical
commercialorhospital
environment
NoteUtisthea.c.mainsvoltagepriortoapplicationofthetestlevel

Related product manuals