Figure 5-17
The procedure for scanning a
panoramic/cephalometric plate is the
same as for the intraorals.
It is possible to select how many
plates to scan, and what size. As
with intraorals, you can choose
between 150, 300, and 600 DPI.
Scan time will be longer as you
choose higher DPI. To switch
between panoramic and cephalo-
metric, click on the “P” or “C” symbol
in the lower right hand corner. To
change between inch and metric,
click “IN” or “CM”.
Choose your imaging plates by click-
ing on each of the imaging plate
boxes you want to acquire and then
clicking on “scan”.
The plates in black will be skipped (not acquired).
To save the template, click on the template menu choice and save.
When the scan is complete, the I/O images will appear on the screen.
Note: if an error occurs during the scan, the scanner will stop. Follow the troubleshooting
guidelines to start scanning again. If the resulting image is not diagnostic, erase the imag-
ing plate and start again.
36
User and Service Manual
Doc # M010-004WWE July 2005
Gendex DenOptix® QST

Other manuals for Gendex DenOptix QST

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gendex DenOptix QST and is the answer not in the manual?

Gendex DenOptix QST Specifications

General IconGeneral
BrandGendex
ModelDenOptix QST
CategoryDental equipment
LanguageEnglish

Related product manuals