Articulated Arm - Horizontal Drifting
032-0207-EN Rev 0
3-6
(Continued)
4. Locate the Friction screw. Using a 4 mm Allen wrench, tighten or loosen
as necessary to get the proper amount of friction or drag to prevent
horizontal drifting.
Figure 3-5 Adjusting the Friction Screw
5. Install the end cap cover.
6. Restore power to the equipment:
FOR A SYSTEM WITH A LINE CORD (120 V ac only), plug in the
line cord.
FOR A HARD-WIRED SYSTEM, reconnect the mains supply
(external power source), and use a meter to verify the mains supply is
connected properly.
7. Turn ON the power switch.