Articulated Arm - Inspection
032-0207-EN Rev 0
3-12
(Continued)
CAUTION
The pins referred to in the following step should never be adjusted in
the field. This information is given to determine if you need to contact
Gendex Technical Support.
4. Ensure that the pins are fully inserted into the casting, and the
pantograph links are seated on the shoulder of the pin. If the pins have
moved outward more than 0.015" (0.38 mm or the thickness of a
business card), contact Gendex Technical Support.
Note! The pins are held in with sealed set screws-NEVER REMOVE THE
SEALANT-this may void the warranty.
5. If the arm is squeaking, lubricate the shoulder of the pins where they
meet the pantograph link with "Tri-Gel" lubricant (Gendex part number
1603-0001).
6. Exercise the arm for several cycles.
7. Check the arm for balance and smooth operation. If drifting is observed,
follow the adjustment procedure that corresponds to the type of drifting.
8. Follow the steps outlined in Closing an End of the Articulated Arm on
page 1-10 to install the trim covers and then restore power.
Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER