032-0207-EN Rev 0
JP1 - Setting for 120 V ac
4-5
(Continued)
4. To enable the neutral line fuse, cut jumper JP1 on the Converter Board
(124-0292G1), as shown in Figure 4-4.
5. Follow the steps outlined in Closing the Master Control on page 1-8 to
close the Master Control
Figure 4-4 Cutting Jumper JP1
Cut
Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER