EasyManua.ls Logo

Gendex Expert DC - Page 51

Gendex Expert DC
62 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
032-0207-EN Rev 0
Logic Board - Test Points
5-17
(Continued)
Table 5-7 Logic Board Test Points Description
Test
Point
Return Description Type Nom. Range
Values
Notes
TP8 TP37, 41 or 42 5V Reference Analog 4.75 to 5.25 V
TP9 TP37, 41 or 42 2.5V Reference Analog 2.4 to 2.6 V
TP10 TP37, 41 or 42 Exposure Switch Digital 0 or 5 V 0 V during exposure
switch closure
TP12 TP37, 41 or 42 X-ray On Digital 0 or 5 V 5 V during exposure
TP14 TP37, 41 or 42 10V Reference Analog 9.5 to 10.5 V
TP16 TP37, 41 or 42 15V Output Analog 14.5 to 15.5 V Power supply
TP17 TP37, 41 or 42 5V Output Analog 4.75 to 5.25 V Power supply
TP18 TP37, 41 or 42 -15V Output Analog -14 to -16 V
TP28 TP37, 41 or 42 Filament I sense Analog 1.9 to 2.3 V During standby;
2 V/ 1A
TP29 TP37, 41 or 42 Preheat I Command Analog -1.9 to -2.3 V During standby
TP30 TP37, 41 or 42 mA Setpoint Analog 6.7 to 7.3 V
TP31 TP37, 41 or 42 mA Feedback Analog 0 V During exposure;
1 V/1 mA
Note: There is no
voltage at this test
point unless an
exposure is being
made
TP32 TP37, 41 or 42 HV I Sense Analog 0V During standby;
1 V/2 A
TP33 TP37, 41 or 42 kV Feedback Analog 0V During exposure;
1 V/20 kV
Note: There is no
voltage at this test
point unless an
exposure is being
made
TP34 TP37, 41 or 42 kV Setpoint Analog 3.1 to 3.4 V
TP37 NA Analog Ground Analog NA
TP39 TP40 5V Digital Analog 4.75 to 5.25 V Power Supply
TP41 NA Analog Ground Analog NA
TP42 NA Analog Ground Analog NA
TP43 TP37, 41 or 42 X-ray Indicate Digital 0 or 5 V 5 V during exposure
Printed on: 20 Apr 2012, 11:34:10 am; Printed by: UWE.ZELLER