Geometrics Inc. G-882 Cesium Marine Magnetometer Page
Here are the current list of commands and the format of each:
Command Format: Description:
----------------- ----------- ----------------------------------
Set Cycle time Byte 1: 'C' Set time in 0.01 sec increments
2: x MS digit of number ('0'-'9')
3: x 3S digit of number
4: x 2S digit of number
5: x LS digit of number
x: x 5 MS optional char ('0' or '5')
6/7: CR carriage return
[Note: the 5 MS char is optional. It was added
to allow setting the cycle time to more precision
after the initial software release]
Set A/D ch's Byte 1: 'A' Enable/disable A/D channels
2: x '0' = turn off channel; '1' = turn on
3: x select channel #('0'-'7')('0'-'7' for CM-
221)
4: x MS digit of counter # ('0' or '1')
5: x LS digit of counter # ('0' - '9')
6: CR carriage return
[Note: characters 4 and 5 can be omitted. If this is done the
command will default to counter 0.]
Change Baud Rate Byte 1: 'B' Baud rate change command
2: x MS char (1,0,0,0,0,0,0)
3: x S char (9,9,4,2,1,0,0)
4: x 3S char (2,6,8,4,2,6,3)
5: x 2S char (0,0,0,0,0,0,0)
6: x LS char (0,0,0,0,0,0,0)
7: CR carriage return