Geometrics Inc. G-882 Cesium Marine Magnetometer Page 86
Symptom
Probable Causes
Corrective Actions
Long warm-up time
Low voltage
Faulty sensor cable
connection
Heater setting too cool
Excessive cold
Defective internal sensor or
electronics components
Increase voltage (minimum 24
VDC at the electronics) or
repair the Coax cable
Connect the fish directly to
the deck cable or cable
adapter
Disconnect sensor from
electronics and carefully
clean the pins and
sockets*
Adjust heater to 34.5 Kohms*
Normal allow extra time for
sensor to warm up in
extremely cold environments
Return fish to Geometrics for
repair
Noisy magnetic field readings
Sensor not oriented correctly
Magnetically “Noisy”
location
Signal amplitude too low
Sensor cable or connector
worn or damaged
Use MagPick IGRF and CsAz
software to model magnetic
field and sensor behavior, then
determine a correct orientation
Move to a magnetically clean
area
See next section on low signal
causes
Replace sensor cable and
connector assembly. Check
connector on electronics
module for wear and replace if
required*
Low Signal Level
Sensor not oriented correctly
Use Magpick IGRF and
CSAZ software to model
magnetic field and sensor