Installation instructions
5018606-00 Rev.C
Edition 0704
Connections
Water consumption
about 15-19 liters/phase
Cold water
Water quality Driinking water with max 5 °dH
Connection 15 (1/2") mm
Pressure 100-800 kPa
Flow rate, min 20 l/min
Hot water
Water quality Preheated drinking water with max 5 °dH
Temperature 45-60 °C (113 - 140 °F)
Connection 15 (1/2") mm
Pressure 100-800 kPa
Flow rate, min. 20 liters/min
Dist./De-ion. water
Temperature, max 90 °C (194 °F)
Connection 20 (3/4") mm
Pressure 50-800 kPa. Where the pressure is lower than 100 kPa, a separate
pressure booster pump must be connected. This is available as an
option. Where a separate feeder pump is used, the pressure must
not exceed 60 kPa (risk of serious damage to the pump).
Flow rate, min 20 liters/min
The maximum conductivity is determined by the desired washing result.
Steam
Temperature, max 160 °C (320 °F)
Connection 15 (1/2") mm
Pressure 300-500 kPa
Consumption 0.5kg/min at 300 kPa
Waste, water
ø 50 mm (2")
Capacity 50 liters/min
Waste, air
ø 63 mm
Capacity 125-150m
3
/h, 35 °C (95 °F)
Humidity 100% 40 sec< 40% after 2 min.