Using a pouch insert
The pouch eater is designed to clean and disinfect pouches lled with liquid. If there is
coagulate in the liquid, the operation of the pouch eater cannot be guaranteed. In such
cases, the result of cleaning and disinfection depends on the size of the coagulate and the
material in the pouch.
Inserting and removing a pouch insert to use the machine for other items
If the pouch eater is not correctly installed, the machine may not
work properly .
Beware of injury by sharp edges.
Removal
To remove the upper part of the pouch insert, press the ange on the quick release
couplings (1) downwards.
Lift out backward and upward.
Installation
Installing is done in the reverse order. Take care to check that the front bar correctly
meets the attachment on the front edge (2). Fit together the coupling and press. A distinct
“click” must be heard.