EasyManua.ls Logo

Getinge 9120 - Rinse2

Getinge 9120
168 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RINSE2 STANDARD
VALUE (type)
PF10 RINSE2 SELECTOR: value 0.1
Rinse 2 selector (Main parameter for selecting or deselecting the whole
rinse 2.)
0 (P)
CT05 RINSE2 FILLTYPE: value 1, 2, 3, 4, 5 (CW, HW, CW+HW, booster, HCW)
Rinse 2 water type selector (parameter for selecting water to be used in
rinse 2
1=cold water, 2=hot water, 3=mixture of cold and hot water, 4=filling from
booster tank, 5=high quality water).
2 (P)
LI10 RINSE2 TEMP SP: value 0-90°C
Rinse 2 temperature set point.
0°C (P)
TI07 RINSE2 TIME: value 0-60 min
Rinse 2 time at selected rinse 2 temperature set point.
1 min (P)
CT14 RINSE2 LOOPS: value 1-3
Rinse 2 number of desired loops 1-3.
1 (P)
PF13 RINSE2 GUARANTEE: value 0.1
Rinse temperature guarantee selector 2. (0=rinse time starts directly
whatever temperature achieved, 1=rinse time starts when selected rinse
temperature is reached)
1 (P)