9122, front panels service
A
A
LOAD END
(LH)
LOAD END
(RH)
* Loading ramp,
(floor mounted)
* Loading ramp,
(floor mounted)
LOAD END
(LH)
FRONT, "LH" LOAD
3003 [9'-10
1
4
"] - Unit height
2850 [9'-4
1
4
"] - Unit height
3200 [10'-6"] - Rec. height (ceiling)
FRONT, Floor "RH"
FRONT, Pit "RH"
2700 [8'-10
1
4
"] - Unit length
2991 [9'-9
3
4
"] - Barrier Trim panels
1225 [4'-
1
4
"] 1225 [4'-
1
4
"]
73 [0'-2
7
8
"]
3020 [9'-10
7
8
"] - Rec. Wall opening
152 [0'-6"]
2870 [9'-5"] - Rec. wall opening
3020 [9'-10
7
8
"] - Rec. wall opening
3350 [10'-11
7
8
"] - Rec. height (ceiling)
200 [0'-7
7
8
"]
Rec. dist.
to wall
Barrier Trim
31 [0'-1
1
4
"]
15 [0'-
5
8
"]
127 [0'-5"] 940 [3'-1"]
(Hook door)
BARRIER TRIM PANELS;
"Hooked door" (US-model)
Not included in machine
1270 [4'-2"]
- Rec. Service access
330 [1'-1"]
1217 [3'-11
7
8
"]
Electrical
Fallback
120 [0'-4
3
4
"]MAX
Rec. wall thicknes
2994 [9'-9
7
8
"] - Barrier Trim panels
100 [0'-3
7
8
"]
970 [3'-2
1
4
"]
BARRIER TRIM PANELS;
"Hinged door"
Not included in machine
Other measurement, see "hooked door"
002940
Page 80 of 92
<Doc_INS><Doc_6001341302><Rel.A><Lang_en>

Other manuals for Getinge 9125

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge 9125 and is the answer not in the manual?

Getinge 9125 Specifications

General IconGeneral
BrandGetinge
Model9125
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals