OPERATING REQUIREMENTS
Surface temp clean and soiled sides <50°C
Humidity level Humidity dropping to less than 40% after approx 5
min drying
Noise level, see note 14 Approx 65 db (A) Inbuilt machine
Radiant heat loss Clean side 2.25 kW
Soiled side 0.90 kW
Service area 5-10 kW
Power consumption Steam heated Max 5.2 KW
Electrically heated max. 65 kW
Air exhaust during drying 800 m³/h
Page 82 of 92