Notes:
1. It is the customer's responsibility to insure by use of pressure regulators or other means that
maximum specified pressures are not exceeded.
2. It is the customer's responsibility to eliminate water or steam hammer conditions should they
occur in the service pipes.
3. It is the customer's responsibility to ensure that water supplies are properly protected against
internal cross connection according to local building and pipe regulations.
4. It is the customer's responsibility to provide a properly sized and located drainage system in
accordance with the relevant national regulations in the area. Consider what other equipment is
connected to drain to eliminate slow drainage or backup. If conditions dictate that the washer-
disinfector will dump water directly into the pit, the pit should be watertight.
5. It is the customer's responsibility to provide a service area with a ventilation system according
to the following specifications:
Alt.1) A system with a constant draught of 1200-1200 m³/h.
Alt 2) A system with a basic draught of 500-800 m³/h and a controlled forced draught during
drying (1200-2000 m
3
/h). (When forced draught ventilation is required, it is signal-activated.)
The system should be a well-functioning, corrosion proof & watertight ventilation system. Direct
the ventilation back towards the washer-disinfector.
6. It is the customer's responsibility to complete all electrical connections using properly sized
wiring in accordance with national regulations.
7. It is the customer's responsibility to provide a fused switch in all electrical supply lines to the
washer-disinfector area.
8. The noise level value applies when the washer-disinfector and service area are built in.
Measured 1 meter in front of the washer and 1.5 m above floor level.
*) The measurement refers to a floorstanding unit. Deduct 150 mm for units installed in a pit
**) Measurements refer to double door installation. Deduct 200 mm for a door.