M
michael81Sep 8, 2025
What to do if there is no LED on my Getinge Medical Equipment?
- Kkristen66Sep 8, 2025
If there is no LED on your Getinge Medical Equipment, the power drive may not be awake. Depress the left hand control trigger once.
What to do if there is no LED on my Getinge Medical Equipment?
If there is no LED on your Getinge Medical Equipment, the power drive may not be awake. Depress the left hand control trigger once.
Why is my Getinge Arjohuntleigh Citadel Plus Medical Equipment bed making an audible tone?
If your Getinge Medical Equipment is emitting an audible tone, it is likely operating from the backup battery. To resolve this, check that the power supply cord is properly plugged in and that the electricity supply is functioning correctly.
What to do if the deck cannot be lowered on Getinge Medical Equipment?
If the Getinge Medical Equipment deck cannot be lowered, it may be due to a height control software error. Raise the deck to its maximum height to reset the software.
What does error code E300 mean on Getinge Arjohuntleigh Citadel Plus?
If you see the error code E300 displayed on your Getinge Medical Equipment, it means a control button has been depressed for more than 90 seconds. Remove pressure from all control buttons. If the error code persists, contact an ArjoHuntleigh approved service agent.
What to do if the deck cannot be lowered and error code AES is displayed on Getinge Arjohuntleigh Citadel Plus?
If the deck of your Getinge Medical Equipment cannot be lowered and the error code AES is displayed, it indicates that the anti-entrapment system has been activated. Remove any obstructions from below the deck. If the bed still cannot be lowered, contact an ArjoHuntleigh approved service agent.
Why does the alarm sound when the patient moves on the Getinge Medical Equipment bed?
If the alarm sounds when a patient moves on the Getinge Medical Equipment bed, the movement detection threshold may be set too low. Increase the VariZone patient movement detection threshold setting.
What does error E102 mean on Getinge Medical Equipment?
If your Getinge Medical Equipment displays error code E102, it means the time out on 10 seconds for the roller (unsteady weight). It may be that the bed was touched or leaned on. Repeat the process with no one touching the bed.
What does error E001 mean on Getinge Arjohuntleigh Citadel Plus?
If your Getinge Medical Equipment displays error code E001, it means the bed was zeroed on battery. Connect the bed to the mains supply and re-initialize the weighing system.
Why will my Getinge Arjohuntleigh Citadel Plus bed not steer?
If your Getinge Medical Equipment bed will not steer, ensure the caster brake pedals are in the neutral position, as the caster steer lock may be engaged.
Why are the TV / lighting controls not working on my Getinge Arjohuntleigh Citadel Plus Medical Equipment?
If the TV/lighting controls are not working on your Getinge Medical Equipment, check to ensure the nurse call cable is of the correct type and properly connected.
Brand | Getinge |
---|---|
Model | Arjohuntleigh Citadel Plus |
Category | Medical Equipment |
Language | English |