CARDIOHELP SETUP GUIDE
4
10 Make sure there is at least 60 cm (2 ft) between the top
of the Cardiohelp and the bottom of the Prime Bag.
2 ./60 cm
11 Remove the sterile protection “stubs” from the lines.
Connect the blue line from the Prime Set to the blue line
of the Table Set. Repeat for the red lines. Leave clamps
on the red line open.