CARDIOHELP SETUP GUIDE
6
ArtVen
--
7.71
4000
$UWHULDOEXEEOHGHWHFWHG
-107
300
326
17 Open the Clamp on Blue Line to passively prime the Set.
Cardiohelp System
Priming the System
18 Once the flow stops, set speed to 3000 rpm for 2 min.
16
5.62
3000
-----
$UWHULDOEXEEOHGHWHFWHG
-27
205
221
20.8
2
min
3,000
rpm
19 Reduce speed to 0 rpm for at least 5 seconds. Reset the
flow/bubble sensor. Tap the tube and the HLS Module
with your hand to remove gas bubbles.
--
-0.04
0
$UWHULDOEXEEOHGHWHFWHG
77
81
80
ArtVen
/30XQWHU*UHQ]H
-0.04
0
$ODUPHLQVWHOOXQJHQ$UW%ODVHQ
20 Set the speed to 4000 rpm for 1 minute and then reduce
speed to 0 rpm. De-air the 3 Luer lock connectors. If a
bubble is detected, go back to step 18.
1
min
4,000
rpm