Servo-i Ventilator System Disassembling and assembling
Revision 02 Service Manual 4 - 7
4
• While holding the label strip in correct position,
remove the protective foils (3 and 4).
• Press the label strip against the adhesive area.
• Remove the protective foil (5).
• Press the front panel film (6) firmly against the front
panel to ensure proper adhesion.
Special functions keys label
• Insert the Special functions keys label strip (7).
• Check its position through the button windows of
the front panel film. Adjust if necessary.
Align the label strip with great care. Once
installed the process cannot be reversed.
• While holding the label strip in correct position,
remove the protective foils (8 and 9).
• Press the label strip against the adhesive area.
• Remove the protective foil (10).
• Press the front panel film (11) firmly against the
front panel to ensure proper adhesion.
@
SVX9036
SVX9037
A
SVX9038
SVX9039
?
SVX9035
=
<
SVX9033
SVX9034
;
9
SVX9031
SVX9032