Disassembling and assembling Servo-i Ventilator System
4 - 8 Service Manual Revision 02
4
Patient Unit
Front cover
To remove the Patient Unit front cover (1):
Remove the screw covers and the screws (2).
Remove the two ventilation covers (3).
Remove the two screws (4).
Carefully lift off the Patient Unit front cover (1).
Note: When assembling the unit, make sure that the
inspiratory and expiratory pressure transducer tubes
and the PC 1785 and expiratory valve coil cables are
not damaged by the Patient Unit front cover (1).
With power supply connected to the Servo-i,
there are energized electrical components inside
the unit. All personnel must exercise extreme
caution if fault tracing or adjustments are
performed with power supply connected and
with the Patient Unit front cover removed.
SVX9041
H6
SVX9042
SVX9043
;9
;
9
SVX9044
6