Servo-i Ventilator System Disassembling and assembling
Revision 02 Service Manual 4 - 9
4
PC 1771, PC 1772 and PC 1784
Remove the Patient Unit front cover.
PC 1771 Control (1):
Release the locks on the PC board guides.
Carefully pull out the PC board.
PC 1772 Monitoring (2):
Disconnect PC 1789 cable connector (3). This is
the cable for the optional PC 1789 Remote alarm
connector.
Release the locks on the PC board guides.
Carefully pull out the PC board.
PC 1784 Expiratory channel (4):
Disconnect pressure transducer tubes (5 and 6)
Disconnect PC 1785 cable connector (7)
Disconnect Exp. valve coil cable connector (8).
Release the locks on the PC board guides.
Carefully pull out the PC board.
Note: When replacing PC 1771 Control, PC 1772
Monitoring or PC 1784 Expiratory Channel, it can be
necessary to re-install the System SW. For further
information refer to section 'Replacing PC boards' in
this chapter.
AC/DC Converter
Remove the Patient Unit front cover.
Disconnect cable connectors (1 and 2).
Remove the screws and nuts (3).
Remove the screw (4).
Carefully lift out the AC/DC Converter (5).
SVX9075
SVX9076
;<
99
H6
SVX9088
?=@;<
96H