Disassembling and assembling Servo-i Ventilator System
4 - 10 Service Manual Revision 02
4
Module unit including PC 1775 Plug-and-
play back-plane
• Remove the Patient Unit front cover.
• Remove the AC/DC Converter.
• Disconnect the nebulizer cable connector (1).
• Remove the screw (2).
• Remove the screws (3).
• Disconnect the Module unit/PC 1775 from PC 1770
Main back-plane at the connector (4).
• Carefully lift out the Module unit (5).
Note: When assembling the unit, make sure that the
spacer (6) is in position with its flat side facing down.
The spacer, designed as a wedge, should be pushed
inwards before the screws (3) are tightened.
PC 1775 Plug-and-play back-plane is mounted on
the Module unit with the screws (7).
SVX9079
?
?
?
?
SVX9078 SVX9077