Servo-i Ventilator System Disassembling and assembling
Revision 02 Service Manual 4 - 11
4
PC 1778 DC/DC & Standard connectors
• Remove the Patient Unit front cover.
• Remove the screws (1).
• Carefully pull out PC 1778 (2).
Internal fan
• Remove the AC/DC Converter.
• Remove PC 1789 Remote alarm connector if
mounted.
• Disconnect the fan cable connector (1).
• Remove the screws (2).
• Carefully lift out the Internal fan (3).
Note: When assembling, make sure that the rubber
seal (4) is mounted as shown in the illustration.