7
Revision 02 Service Manual 7 - 3
Servo-i Ventilator System Preventive maintenance
Preventive maintenance
Maintenance kit, 5,000 hours
Only original parts from the manufacturer must be
used. Spare parts and maintenance kits can be
ordered from your local Maquet representative.
When performing this maintenance, a 'Maintenance
kit, 5,000 hours' should be used.
The following parts shall be replaced and they are
included in the 'Maintenance kit, 5,000 hours':
A. Filters for the gas modules
B. Nozzle units for the gas modules
C. Bacteria filter for the inspiratory pressure
transducer
D. Bacteria filter for the O
2
cell
Performing the Preventive maintenance
Disassembling and assembling of the unit is
required when replacing parts included in the
'Maintenance kit, 5,000 hours'. If not stated
otherwise, refer to chapter 'Disassembling and
assembling' for instructions.
The letters A – D in the text below refers to the
description of the Maintenance kit above.
Prepare the unit as described in section
'Preparations' above in this chapter.
Gas modules
A. Replace filters in the gas modules. When
replacing filter, move the rubber seal from the old
to the new filter.
B
A
D
C
SVX9066
A