2 INSTALLATION
Note! Be careful when lifting the machine manually.
Weight 92 kg.
The floor surface where the machine is to be placed must be
plane within ± 5 mm.
Take the cardboard box off the machine.
Place the machine in the desired location, and use the feet
adjustment screws so that the machine is standing stable.
The adjusting feet must not be unscrewed more than 50 mm
from the underside of the machine. The recommended
amount of space in front of the machine is 1000 mm.
Use a level to check that the machine is level within ±2 mm.
Use the level to check on the door and on the guides.
When the machine is balanced, adjust the feet so that the
lock catch meets the middle of the lock hole and so that the
door closes properly.
Ensure that there is an air gap on the back of the machine for
exhaust from the condenser.
000456
Figure 1. Measure when level
2 INSTALLATION
10 (48) INSTALLATION MANUAL GETINGE WD15 CLARO