LPS10
- 18 -
V1.0
Holding Registers:
Address Description Format
0
RS485 Modbus-RTU Baud Rate:
0=9600
1=19200 (default)
2=38400
3=57600
4=115200
16-bit Integer
1
RS485 Modbus-RTU parity and stop bits:
0=8N1
1=8N2
2=8E1 (default)
3=8E2
4=8O1
5=8O2
16-bit Integer
2 Modbus-RTU address (1…247, default=1). 16-bit Integer
5
Temperature unit of measurement:
0=°C (default)
1=°F
2=K
16-bit Integer
6
Operating time alarm threshold from first power on (0…65535
days, default=730 days 2 years)
16-bit Integer
7
Operating time alarm threshold from last power on (0…65535
days, default=730 days 2 years)
16-bit Integer
8
Internal temperature alarm threshold in the set unit of meas-
urement (0…150 °C, default=80 °C)
16-bit Integer
9
Internal relative humidity alarm threshold in % (0…100 %,
default=25%)
16-bit Integer
10
Internal pressure alarm threshold in hPa (800…1100 hPa, de-
fault=1100 hPa)
16-bit Integer
All alarm thresholds are integer values with resolution 1 (1 day, 1 °C/°F/K, 1 %RH, 1
hPa).

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GHM Delta OHM PYRAsense LPS10 Series and is the answer not in the manual?

GHM Delta OHM PYRAsense LPS10 Series Specifications

General IconGeneral
BrandGHM
ModelDelta OHM PYRAsense LPS10 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals