LPS10
- 19 -
V1.0
8 MEASUREMENT WITH ANALOG OUTPUT (only LPS10MA…)
Calculation of the global irradiance E
e
(W/m
2
) as a function of the analog output value
V
out
(in V) or I
out
(mA).
Analog output
Global irradiance
0…20 mA
E
e
= E
is
+ [(E
fs
– E
is
) I
out
/20]
4…20 mA
E
e
= E
is
+ [(E
fs
– E
is
) (I
out
4)/16]
0…1 V
E
e
= E
is
+ [(E
fs
– E
is
) V
out
]
0…5 V
E
e
= E
is
+ [(E
fs
– E
is
) V
out
/5]
0…10 V
E
e
= E
is
+ [(E
fs
– E
is
) V
out
/10]
Where E
is
and E
fs
are respectively the global irradiance (in W/m
2
) corresponding to
the analog output initial and full scale.
Considering the default values E
is
= 0 and E
fs
= 2000 W/m
2
, the previous formulas
can be simplified into the following ones:
Analog output
Global irradiance
0…20 mA
E
e
= 100 I
out
4…20 mA
E
e
= 125 (I
out
4)
0…1 V
E
e
= 2000 V
out
0…5 V
E
e
= 400 V
out
0…10 V
E
e
= 200 V
out

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GHM Delta OHM PYRAsense LPS10 Series and is the answer not in the manual?

GHM Delta OHM PYRAsense LPS10 Series Specifications

General IconGeneral
BrandGHM
ModelDelta OHM PYRAsense LPS10 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals