LPS10
- 5 -
V1.0
Technical Specifications According to ISO 9060:2018
Classification Spectrally Flat Class A
Response time (95%) < 2 s
Zero offset
a) response to a 200 W/m
2
thermal
radiation
<±7W/m
2
b) response to a 5 K/h change in
ambient temperature
<±2W/m
2
c) total zero offset including the effects
a), b) and other sources
<±10W/m
2
Long-term instability (1 year) <±0.5 %
Non-linearity <±0.2 %
Directional response
(up to 80° with 1000 W/m
2
beam)
<±10 W/m
2
Spectral error <±0.2 %
Temperature response (-10…+40°C) <±0.5 %
Tilt response <±0.2 %
Diagnostic sensors
Internal Temperature
Measuring range -40…+80 °C
Resolution 0.1 °C
Accuracy ±0.5 °C (0…60 °C)
Internal Relative Humidity
Measuring range 0…100%
Resolution 0.1%
Accuracy ±3% @ 25 °C (20…80%RH)
Internal pressure
Measuring range 300…1100 hPa
Resolution 0.1 hPa
Accuracy ±1 hPa (0…60 °C)
Tilt sensor (only LPS03MxT)
Measuring range -180°…+180°
Resolution 0.1°
Accuracy < 0.5°

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GHM Delta OHM PYRAsense LPS10 Series and is the answer not in the manual?

GHM Delta OHM PYRAsense LPS10 Series Specifications

General IconGeneral
BrandGHM
ModelDelta OHM PYRAsense LPS10 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals