LPS10
- 5 -
V1.0
Technical Specifications According to ISO 9060:2018
Classification Spectrally Flat Class A
Response time (95%) < 2 s
Zero offset
a) response to a 200 W/m
2
thermal
radiation
<±7W/m
2
b) response to a 5 K/h change in
ambient temperature
<±2W/m
2
c) total zero offset including the effects
a), b) and other sources
<±10W/m
2
Long-term instability (1 year) <±0.5 %
Non-linearity <±0.2 %
Directional response
(up to 80° with 1000 W/m
2
beam)
<±10 W/m
2
Spectral error <±0.2 %
Temperature response (-10…+40°C) <±0.5 %
Tilt response <±0.2 %
Diagnostic sensors
Internal Temperature
Measuring range -40…+80 °C
Resolution 0.1 °C
Accuracy ±0.5 °C (0…60 °C)
Internal Relative Humidity
Measuring range 0…100%
Resolution 0.1%
Accuracy ±3% @ 25 °C (20…80%RH)
Internal pressure
Measuring range 300…1100 hPa
Resolution 0.1 hPa
Accuracy ±1 hPa (0…60 °C)
Tilt sensor (only LPS03MxT)
Measuring range -180°…+180°
Resolution 0.1°
Accuracy < 0.5°